EEG & EMG HKT88 3200 Machine

HKT88-3200 Digital Brain Electric Activity Mapping collects EEG signal with electrodes, via integrated amplification, A/D transformation, PC auto-analysis, FFT, to form electroencephalogram that displays with color depth. The product is applicable for checking such diseases as epilepsy, intracranial inflammation, cerebrovascular diseases and brain tumors.

Performance
  • 32 channels of EEG
  • Sampling rate: 200 dots/s
  • Accuracy: 12bit
  • Input impedance: ≥10MΩ
  • Patient leak current: < 10µA
  • Noise level: ≤5µVp-p
  • CMRR: ≥90dB
  • Magnification multiple: 10000
  • Filter constant: all digital and free enactment
  • Display speed(paper speed): 5, 10, 15, 30, 60, 120 mm/s
  • Amplitude: 1, 1.5, 2, 3, 5, 7.5, 10, 12, 15, 20, 30, 50 mm/50µV
  • Playback speed:1 time, 2 times, 3 times, 10 times, 20 times, 40 times, 60 times
  • 50Hz interference suppression: ≥30dB
  • Safety type:Class II, type BF applied part
Compare
Category:

Description

Product parameter
Dimension: 186 mm(L)×186 mm(W)×89.2 mm(H)
Weight: 750g
Operational environment:
Temperature: +5℃~+40℃
Relative humidity: 25%~95% (no condensation)
Power supply: USB 5V DC
Atmospheric pressure: 700hPa~1060hPa
Storage environment:
Temperature:-20℃~+55℃
Relative humidity: ≤90%
Atmospheric pressure: 700hPa~1060hPa