Transport Infant Incubator HT-4000

SPECIFICATION:

  • Air mode and Baby mode controlled by micro-computer;
  • Use AC Power and DC Power alternatively, DC 12V or DC 24V Power on the ambulance;
  • Set temperature, air temperature, baby temperature and internal battery voltage can be displayed separately;
  • Double wall Acrylic glass hood with side door so that the bassinet can be pulled out from it;
    Natural air flow humidity;
  • The height of whole unit can be adjusted;
  • With Oxygen cylinder and Oxygen supply configuration;
  • Observation lamp

Standard configuration:

  • Main body (including the Acrylic glass hood, temperature controller, bassinet, internal battery, and observation lamp)
  • Oxygen cylinders
  • Oxygen supply system
  • Skin temperature sensor
  • I.V. Pole, mattress
  • Adjustable stand
Compare

Description

Mode of control Air mode and baby mode controlled by micro-computer
Control range of air temperature 25°C-37°C(Override mode 37-38°C)
Control range of baby temperature 34°C-37°C(Override mode 37-37.5°C)
Accuracy of skin temperature sensor ≤0.3°C
Variability of temperature ≤1.0°C
Uniformity of temperature ≤1.5°C
Noise inside hood ≤60dB(A)
Failure alarm Over-temperature alarm
Temperature deviation alarm
Sensor alarm
Motor Fan Failed Alarm
Power failure Alarm and so on
Running time of internal battery 90min (1 internal battery)
AC Power supply AC220V-230V, 50Hz
DC Power supply DC12V/10A or DC24V/6A
Input power 400VA
Main Body Weight 80kg
Dimension 109*65.5*76cm
Cabinet Weight 45kg
Dimension 138*61*40cm